Code N Performance
|
Code C Performance
|
Torque Range
|
Factory Set @ Transducer Full Scale Torque. Field Resettable to any lower value; see Note 1.
|
Units of Measure
|
User may select from lbf-in, lbf-ft, ozf-in, ozf-ft, N-m, kN-m, N-cm, kgf-m, kgf-cm, gf-cm without re-calibration
|
Accuracy2 (nonlinearity, hysteresis & repeatability)
|
<±0.1 (End Point, % of Range)
|
<±0.05 (End Point, % of Range)
|
Bi-polar Shunt Calibration Enable
|
From Stator Switches (one CW, one CCW), via TTL I/O, or PC Com Port using furnished software.
|
Zero Drift (% of Range/deg. F.)
|
<±0.002
|
<±0.001
|
Span Drift (% of Reading/deg. F.)
|
<±0.002
|
<±0.001
|
Temperature Ranges (deg. F.)
|
Compensated Range: +75 to +175; Usable Range: -25 to +185; Storage Range: -65 to +225
|
Rotor to Stator Maximum Misalignment (inches)
|
Axial: ±0.5, Radial: 0.3; If Speed Pickup Option is installed then maximums may be reduced - call factory.
|
Rotational Effects (% of Range)
|
#±0.01
|
Analog Output Signals6, Auto-Scaled
|
|
MCRT® 86000V Full Scale Torque3
|
Default is ±10V with 50% overrange (±15V). User may select ±5V with 50% overrange (±7.5V). Caution: see Note 7.
|
MCRT® 87000V Full Scale Torque3
|
Default is ±5V with 300% overrange (±15V). User may select ±10V with 50% overrange (±15V). Caution: see Note 7.
|
Signal Filter Cutoff Frequency4
|
Field selectable from 1 to 1 kHz in ten 1-2-5 steps plus 3 kHz selected from a remote PC using software furnished.
Filters have Bessel Response; no delay distortion or overshoot errors.
|
Frequency Modulated Output, Auto-Scaled
|
Frequency: 10±5 kHz or 20±10 kHz or 40±20 kHz; field changeable (Default = 10±5 kHz); TTL square wave output.
|
Overrange (% of Range, see Application Note 20805)
|
50 to 300; model/user selection dependent - see above. Applies to analog, Digital and FM (except for negative side) Outputs.
|
System Resolution (% of Range)
|
0.01
|
Output Noise, All Outputs (% of Range)
|
<0.01 at 1Hz, <0.01 at 10 Hz, < 0.015 at 100 Hz, <0.028 at 1 kHz and <0.041 at 3 kHz.
|
System Response
|
Torque is sampled @ 20 kHz. Its’ 3 db bandwidth is 3 kHz but can be reduced by filters (see above)
|
Rotor-to-Stator Transfer Rate
|
1.25 MBaud
|
RS232, RS422, RS485 Communications Port
|
Com port outputs Torque and Temperature with units of measure. Inputs torque range if other than sensor full scale, selects units of measure,
selects BAUD Rate, and permits remote computer control of the test.
|
BAUD Rate
|
9,600 to 230,400 (default is 115,200); Drivers are protected for short circuit (current limit) and ±15kV ESD protected
|
Maximum Cable Length
|
RS232 = 50 feet, RS422/485 = 4,000 feet; 120 Ohm termination may be accessed via software.
|
Interface Software
|
Provided to interface with Windows-based PC. Includes interconnect cable for a PC.
|
I/O Lines
|
5 input and 2 output, all are user assignable. Default for 5 inputs are +CAL, -CAL, TARE, CLEAR TARE, and RESET MAX/MIN.
|
Status LED’s (on Stator Keypad)
|
Three Color Coded LED’s: Power (Yellow = Power-up, Green = OK, Red = Fault); Data (Green = OK, Red = Data Error);
Rotor Temperature (Green = In Operating Range, Red = Out of Operating Range).
|
Keypad Control Switches
|
+ CAL invokes CW Rotor Shunt Cal, -CAL invokes CCW Rotor Shunt Cal, Both held simultaneously for 5 seconds invokes TARE.
|
Rotor Temperature, Auxiliary Measurement
|
Rotor temperature is outpout via Comp Port. Range is 0 to 185 degrees F.; Accuracy is ±2 degrees. See above for status LED’s.
|
Option Code Z
|
30 to 180 pulses/revolution (model dependent) Zero Velocity Speed Pickup.
|
Supply Voltage/Power5
|
10 to 26 VDC @ 6 Watts with Antenna aligned to 9 Watts with maximum specified misalignments.
|